When completing process oriented tasks, checklists are important part of the process. A thorough explanation of the benefits of checklists can be fount in The Checklist Manifesto

There are many crucial steps in collecting “routine” motion capture data.

The following checklist is a starting point from which Motion Capture Laboratories can build their own.

We encourage users to contribute to this list by sending additional tasks to support@c-motion.com

Before Patient's Arrival

  • Turn on ALL systems
  • Item Verify Position and Orientation
    • Verify the position and orientation of Motion Capture Cameras
    • Verify the position and orientation of Force Platforms
    • Verify the position and orientation of Force Structures
    • Verify the position and orientation of Obstacles or Environment objects
  • Perform System Calibration
    • Perform static and/or dynamic calibrations of data collection devices
    • Verify results against the Standards developed/referenced by your laboratory
  • Verify the protocol being used for the session
    • Verify the experimental protocol for the current data collection session
  • Set appropriate system settings
    • As per experimental protocol, set and verify system settings
      • e.g. sampling rates (synchronise), marker thresholds, ranges, triggers etc.
  • Zero Force Plates and Perform CalTester test
    • Perform Caltester test at a minimum of 2 locations on each force platform
    • Verify against prior laboratory results
  • Prepare necessary markers and supplies

Patient's Arrival

  • Obtain informed consent
  • Record subject details
    • Insure that there are no contraindications for the patient completing the data collection
    • This may include data collection done via questionnaires or measurements
    • If data collection includes patient questionnaires check to make sure that they have been completed correctly
  • Apply markers as per experimental protocol
    • Take a set of digital pictures of the subject for post-hoc review of the marker placement.
  • Record static calibration trial
    • Ensure all markers are visible
  • Verify recorded marker set
    • Verify recorded marker set by launching Visual3D, creating model, and applying model template appropriate for the protocol
  • Remove anatomical markers
    • If necessary, remove anatomical markers not needed for tracking in motion trials
  • Verify tracking using Real time
    • If you switch to the real time mode in Visual3D, you can verify that the tracking markers assigned to each segment are appropriate, and that the force vectors make intuitive sense
  • Record dynamic trials
  • Verify force platform assignments
    • Load data into Visual3D, assign movement trials to model, and verify force platform assignments
    • Remove assignments to partial force contacts
    • Remove assignments when two segments are in contact with the same platform.
  • Execute pipeline for creating event labels.
    • Verify that these event labels make intuitive sense. If not, data collection should be repeated

After Patient Leaves

  • Clean up work station
  • Execute pipeline for processing
  • Load Report Template
  • Verify the report.